VOC Environmental Test Chamber (Sampling Bag Method) fully meets the requirements for laboratory assessment in the automotive industry and interlaboratory comparison tests for VOC detection using the sampling bag method. Automotive interior materials or products are placed in the test chamber or sampling bag, and appropriate methods are used to analyze gas samples, identify the main pollutants and changes in their concentrations over time, and then determine the emission rates and emission characteristics of automotive interior materials and products based on the measured concentrations.
VOC Environmental Test Chamber (Sampling Bag Method) fully meets the requirements for laboratory assessment in the automotive industry and interlaboratory comparison tests for VOC detection using the sampling bag method. Automotive interior materials or products are placed in the test chamber or sampling bag, and appropriate methods are used to analyze gas samples, identify the main pollutants and changes in their concentrations over time, and then determine the emission rates and emission characteristics of automotive interior materials and products based on the measured concentrations.
|
Model |
CZ-V-WIR-24 |
|
Test Chamber Dimensions W×H×D (mm) |
3000×2000×4000=24m³ (Customizable according to customer site conditions) |
|
Door Dimensions W×H (mm) |
2000x1800 |
|
Temperature Range |
RT+15℃ ~ +100℃ |
|
Chamber Air Pressure |
5-50Pa |
|
Fresh Air Exchange Rate |
2~10 times/h |
|
Top Solar Irradiance System |
Infrared, wavelength 280–3000 nm, irradiance 350–1200 W/m², area: W6500 × H3500, meets JAMA / PV3938 standards |
|
Background Concentration |
1. Benzene content <0.015; 2. Toluene ≤0.015; 3. Xylene ≤0.015; 4. Ethylbenzene ≤0.015; 5. Styrene ≤0.015; |
|
6. Formaldehyde ≤0.015; 7. Acetaldehyde ≤0.015; 8. Acrolein ≤0.015; 9. TVOC content <0.350; |
|
Model |
CZ-V-WIR-24 |
|
Test Chamber Dimensions W×H×D (mm) |
3000×2000×4000=24m³ (Customizable according to customer site conditions) |
|
Door Dimensions W×H (mm) |
2000x1800 |
|
Temperature Range |
RT+15℃ ~ +100℃ |
|
Chamber Air Pressure |
5-50Pa |
|
Fresh Air Exchange Rate |
2~10 times/h |
|
Top Solar Irradiance System |
Infrared, wavelength 280–3000 nm, irradiance 350–1200 W/m², area: W6500 × H3500, meets JAMA / PV3938 standards |
|
Background Concentration |
1. Benzene content <0.015; 2. Toluene ≤0.015; 3. Xylene ≤0.015; 4. Ethylbenzene ≤0.015; 5. Styrene ≤0.015; |
|
6. Formaldehyde ≤0.015; 7. Acetaldehyde ≤0.015; 8. Acrolein ≤0.015; 9. TVOC content <0.350; |